Breast irradiation with three conformal photon fields for patients with high lung involvement |
| |
Authors: | Cozzi Luca Fogliata Antonella Nicolini Giorgia Rancati Tiziana Bernier Jacques |
| |
Affiliation: | Medical Physics Unit, Department of Physics Oncology, Institute of Southern Switzerland, University of Milan, Bellinzona, Switzerland. lucozzi@iosi.ch |
| |
Abstract: | Since 2001, 50 breast cancer patients, for whom extensive lung/heart involvement was expected from the use of conventional tangential 2-field technique (2F) owing to complex anatomies, were irradiated using a 3-field conformal technique (3F). Dose plans were designed for both 3F and 2F and a dose volume histogram analysis on ipsilateral lung, heart, and target was conducted. The 3F technique allowed a significant reduction in ipsilateral lung irradiation: mean dose dropped from 16.0±3.8 (2F) to 12.0±2.7 Gy (3F) and V45Gy from 20.7±6.8 (2F) to 3.2±2.9% (3F). Similarly, in patients irradiated to the left breast, heart mean dose was reduced from 8.1 Gy (2F) to 6.8 Gy (3F) and D15% from 19.0 Gy to 14.0 Gy. All differences reached a high degree of significance. The target coverage was not clinically compromised since the slight reduction using 3F compared with 2F is limited to V95% while V90% difference, even if statistically significant, is small: 98.2±1.8% (3F) and 98.8±1.6 (2F). A preliminary report on clinical follow up is also included; with a mean follow up of 15.8 months, no pulmonary or cardiac complications were observed. |
| |