热牙胶充填系统携热头工作状态下的实际温度
引用本文:渠薇,白伟,梁宇红,高学军.热牙胶充填系统携热头工作状态下的实际温度[J].北京大学学报(医学版),2015,47(5):834-837.
作者姓名:渠薇  白伟  梁宇红  高学军
作者单位:(1.北京大学口腔医学院·口腔医院牙体牙髓科,北京100081; 2.北京大学口腔医学院·口腔医院口腔材料研究室,北京100081;3.北京大学国际医院口腔医学科,北京102206)
摘    要:目的:测量热牙胶充填系统携热头工作状态下的实际温度,为临床研究热损伤、观察温度升高对充填材料影响提供理论依据。方法:使用双通道K型电偶温度计测量3种热牙胶充填系统的不同大小携热头和携热头不同部位(工作端尖端和距尖端2 mm、5 mm、10 mm处)的实际温度(设置温度200 ℃),重复测量5次,比较不同充填系统、不同携热头大小、不同测量位点的实际工作温度和设置温度的差异,并用单因素方差分析法分析各因素的影响。结果:BeeFill 40/0.03携热头尖端工作温度最高(198.7±7.7)℃,与设定温度200 ℃差异无统计学意义(P>0.05),其余携热头工作温度均低于设定温度200 ℃,差异有统计学意义(P<0.05),BeeFill 60/0.06携热头距尖端10 mm处温度最低(69.9±4.0) ℃。所有携热头尖端(0 mm)及尖端2 mm处温度较高(112.1~198.7 ℃,中位数140.8 ℃)。结论:大部分热牙胶充填系统的携热头实际工作温度低于设定温度200 ℃。

关 键 词:,温度,根管充填材料,携热头,

Real temperature of the continuous-wave pluggers
QU Wei,BAI Wei,LIANG Yu-hong,GAO Xue-jun.Real temperature of the continuous-wave pluggers[J].Journal of Peking University:Health Sciences,2015,47(5):834-837.
Authors:QU Wei  BAI Wei  LIANG Yu-hong  GAO Xue-jun
Institution:(1.Department of Cariology and Endodontology, 2.Dental Material Laboratory, Peking University School and Hospital of Stomatology, Beijing 100081, China; 3. Dental Clinic, Peking University International Hospital, Beijing 102206, China)
Abstract:Objective:To measure the real temperatures on the pluggers of three continuous-wave de-vices, and to provide theoretical reference to evaluate thermal damage and heat ’s influence on the filling materials.Methods:The dual channel K type thermocouple was contacted to various sizes ’ pluggers in three different continuous-wave devices (BeeFill, Elements, B&L), and the highest temperatures at dif-ferent points ( tip, and 2 mm, 5 mm, 10 mm from the tip) of the pluggers ( preset temperature was 200℃) were recorded.The measurements were performed 5 times.T-test was used to compare the real tem-peratures at the tips with that set on the display and one-way ANOVA was used to compare the tempera-tures of the pluggers in different devices , sizes and points .Results:The highest temperature was at the tip of BeeFill 40/0.03 plugger (198.7 ±7.7) ℃, but there was on statistical differences between that and the preset temperature 200℃.The temperatures of the remaining pluggers were obviously lower than 200 ℃(P<0.05).The lowest temperature of the pluggers was detected at 10 mm from the tip of BeeFill 60/0.06 plugger (69.9 ±4.0) ℃.The highest temperature of each plugger was detected at the tip or 2 mm from the tip (112.1 to 198.7℃,and the median was 140.8℃).Conclusion:The real temperature of most continuous-wave pluggers included in this study is below the set temperature 200 ℃.
Keywords:Temperature  Root canal filling materiak  Plugger