首页 | 本学科首页   官方微博 | 高级检索  
     


Performance of VITEK-2 Compact and overnight MicroScan panels for direct identification and susceptibility testing of Gram-negative bacilli from positive FAN BacT/ALERT blood culture bottles
Authors:M.D. Quesada  M. Giménez  S. Molinos  G. Fernández  M.D. Sánchez  R. Rivelo  A. Ramírez  G. Banqué  V. Ausina
Affiliation:Department of Microbiology, Hospital Universitari Germans Trias i Pujol, Universitat Autònoma de Barcelona, C/Canyet s/n, Badalona, Spain
Abstract:We describe the reliability of the VITEK-2 Compact and overnight MicroScan panels for direct identification and susceptibility testing from the BacT/ALERT blood culture system when using FAN (FA and FN) bottles. A simple procedure, in two centrifugation steps, was designed to remove the charcoal particles present in FA and FN bottles. A total of 113 positive blood cultures showing Gram-negative rods were investigated. Enterobacteriaceae were isolated in 104 cases, and Pseudomonas aeruginosa in nine. The MicroScan system correctly identified 106 (93.8%) of the 113 isolates. The seven identification errors included P. aeruginosa (three), Enterobacter cloacae (one), Escherichia coli (one), Klebsiella oxytoca (one), and Klebsiella pneumoniae (one). The VITEK-2 system correctly identified 109 (96.5%) of the 113 samples obtained directly from the blood culture bottles. The four unidentified isolates were Enterobacter cloacae (two), Escherichia coli (one), and P. aeruginosa (one). MicroScan yielded 4/779 (0.5%) very major errors and 28/2825 (0.9%) minor errors. VITEK-2 yielded 2/550 (0.36%) very major errors, 1/1718 (0.05%) major error, and 32/2373 (1.3%) minor errors. Both systems provided excellent identification (correlation of >90%) and susceptibility (correlation of >98%) results. The average times required to obtain identification and susceptibility results using the direct test applied to the VITEK-2 Compact system were 4.57 ± 1.37 h and 6.52 ± 1.64 h, respectively. The VITEK-2 compact system provided results on the same day that the blood culture was found to be positive.
Keywords:BacT/ALERT  blood culture  FAN bottles  MicroScan  Vitek
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号