Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice
Authors:Wouter J. H. Veldkamp  Lucia J. M. Kroft  Mireille V. Boot  Bart J. A. Mertens  Jacob Geleijns
Affiliation:(1) Department of Radiology, C2S, Leiden University Medical Center, Albinusdreef 2, 2333 ZA Leiden, The Netherlands;(2) Department of Medical Statistics, C2S, Leiden University Medical Center, Albinusdreef 2, 2333 ZA Leiden, The Netherlands;(3) Department of Radiology, C2S, Leiden University Medical Center, PO Box 9600, 2300 RC Leiden, The Netherlands
Abstract:
The purpose of this study was to assess contrast-detail performance and effective dose of eight different digital chest radiography systems. Digital chest radiography systems from different manufacturers were included: one storage phosphor system, one selenium-coated drum system, and six direct readout systems including four thin-film transistor (TFT) systems and two charge-coupled device (CCD) systems. For measuring image quality, a contrast-detail test object was used in combination with a phantom that simulates the primary and scatter transmission through lung fields (LucAl). Six observers judged phantom images of each modality by soft-copy reading in a four-alternative-forced-choice experiment. The entrance dose was also measured, and the effective dose was calculated for an average patient. Contrast-detail curves were constructed from the observer data. The blocked two-way ANOVA test was used for statistical analysis. Significant difference in contrast-detail performance was found between the systems. Best contrast-detail performance was shown by a CCD system with slot-scan technology, and the selenium-coated drum system was compared to the other six systems (p values le0.003). Calculated effective dose varied between 0.010 mSv and 0.032 mSv. Significant differences in contrast-detail performance and effective dose levels were found between different digital chest radiography systems in clinical practice.
Keywords:Digital radiography  Flat-panel detector  Chest radiography
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号