首页 | 本学科首页   官方微博 | 高级检索  
     


Feasibility of extended-coverage perfusion and dynamic computer tomography (CT) angiography using toggling-table technique on 64-slice CT
Authors:Horiguchi Jun  Kiura Yoshihiro  Tanaka Junpei  Fukuda Hiroshi  Kiguchi Masao  Fujioka Chikako  Kurisu Kaoru  Awai Kazuo
Affiliation:aDepartment of Radiology, Hiroshima Kyoritsu Hospital, 2-19-6, Nakasu, Asaminami-ku, Hiroshima, 731-0121, Japan;bDepartment of Neurosurgery, Graduate School of Biomedical Sciences, Hiroshima University, 1-2-3, Kasumi-cho, Minami-ku, Hiroshima, 734-8551, Japan;cDepartment of Clinical Radiology, Hiroshima University Hospital, 1-2-3, Kasumi-cho, Minami-ku, Hiroshima, 734-8551, Japan;dDepartment of Radiology, Division of Medical Intelligence and Informatics, Programs for Applied Biomedicine, Graduate School of Biomedical Sciences, Hiroshima University, 1-2-3, Kasumi-cho, Minami-ku, Hiroshima, 734-8551, Japan
Abstract:

Purpose

The major drawbacks of brain computed tomography (CT) perfusion are limited coverage in the z-axis, radiation exposure and the use of contrast medium, all of which increase when CT angiography (CTA) is performed with double acquisitions. The purpose of this study was to investigate the feasibility of the ‘toggling-table’ technique using a 40-mm detector.

Materials and methods

The clinical usefulness of CT perfusion and 4D-CTA (time-resolved 3D-CTA), as well as the time taken and radiation exposure, were assessed in 14 non-ischemic patients clinically indicated for CT perfusion and CTA.

Results

A perfusion map and 4D-CTA was successfully achieved in all patients. The total time needed for scanning and processing was approximately 30 min per examination. The 80-mm CT perfusion coverage and 4D-CTA images were advantageous for analyses of the anatomy and widely distributed lesions. The estimated effective radiation dose was 2.22 mSv, and the amount of contrast media was 40 mL.

Conclusion

The ‘toggling-table’ technique acts as a ‘one-stop-shop’ protocol for perfusion mapping and 4D-CTA over a wide area, following a single contrast injection and scan.
Keywords:Perfusion   4-D CTA   CT angiography
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号