Clinical MR imaging with a Helmholtz-type surface coil |
| |
Authors: | Reiman, TH Heiken, JP Totty, WG Lee, JK |
| |
Affiliation: | Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, MO 63110. |
| |
Abstract: | Limited-field-of-view radio-frequency receiver antennas provide improved near-field sensitivity for magnetic resonance imaging by decreasing the antenna volume. The Helmholtz-type surface coil, consisting of two flat rings, is an organ-encompassing antenna that takes advantage of this principle to yield an improved signal-to-noise ratio (S/N). The coil was tested in a group of 50 patients and 16 healthy volunteers. Images obtained with the Helmholtz coil demonstrated quantitatively superior S/N of 2.2-fold or greater than that of comparison body coil images, as well as qualitatively superior anatomic resolution. |
| |
Keywords: | |
|
| 点击此处可从《Radiology》浏览原始摘要信息 |
|
点击此处可从《Radiology》下载全文 |
|