Optimum selection of an implantable secondary battery for an artificial heart by examination of the cycle life test |
| |
Authors: | Okamoto Eiji Watanabe Kazuya Hashiba Kunihiro Inoue Taku Iwazawa Eichi Momoi Masato Hashimoto Takuya Mitamura Yoshinori |
| |
Affiliation: | Department of Information Science, School of Engineering, Hokkaido University, Sapporo, Japan. |
| |
Abstract: | An implantable secondary battery is one of the key components in a total artificial heart system. Because a 2 year cycle life is required, the cycle life of the secondary battery as well as its charge and discharge properties are important parameters for selection of an appropriate battery. We carried out cycle life tests on four kinds of rechargeable batteries (a Ni-MH secondary battery, a Ni-Cd secondary battery, a Li-ion battery with a graphite anode, and a Li-ion battery with a nongraphitizable carbon electrode) to determine their suitability as implanted back-up batteries. Each of the batteries was charge/discharge cycled at 37 degrees C to 39 degrees C using a charge current of 1 C ampere, and they were each fully discharged under either pulsatile discharge loads, which mimicked pulsatile operation, or a nonpulsatile load equivalent to the average of the pulsatile loads. The two Li-ion batteries made by different manufacturers both met the minimum requirement of cycle life of more than 1,500 cycles, considering safety coefficient regardless of the discharge pattern. In addition, the temperature increase of these Li-ion batteries (3 degrees C) was lower than that of Ni-Cd and Ni-MH batteries (15-25 degrees C). Out of these four batteries, the two Li-ion batteries are the most suitable for use in a totally implantable artificial heart system. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|