首页 | 官方网站   微博 | 高级检索  
     


Prevalence and accuracy of home sphygmomanometers in an urban population.
Authors:L P Hahn  A R Folsom  J M Sprafka  and R J Prineas
Affiliation:Division of Epidemiology, University of Minnesota School of Public Health, Minneapolis 55455.
Abstract:The prevalence of ownership (7.5 per cent) and the accuracy of home sphygmomanometers were determined in a population-based survey in the Minneapolis-St. Paul metropolitan area. Sixty-four per cent of home sphygmomanometers were accurate within +/- 2 mm Hg of a calibrating sphygmomanometer; another 26 per cent were within +/- 3-6 mm Hg. These results suggest that although many home sphygmomanometers are accurate, some are very inaccurate. Health care providers should advise regular calibration when home sphygmomanometers are used for therapeutic self-management of hypertension.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23