Prevalence and accuracy of home sphygmomanometers in an urban population. |
| |
Authors: | L P Hahn A R Folsom J M Sprafka and R J Prineas |
| |
Affiliation: | Division of Epidemiology, University of Minnesota School of Public Health, Minneapolis 55455. |
| |
Abstract: | The prevalence of ownership (7.5 per cent) and the accuracy of home sphygmomanometers were determined in a population-based survey in the Minneapolis-St. Paul metropolitan area. Sixty-four per cent of home sphygmomanometers were accurate within +/- 2 mm Hg of a calibrating sphygmomanometer; another 26 per cent were within +/- 3-6 mm Hg. These results suggest that although many home sphygmomanometers are accurate, some are very inaccurate. Health care providers should advise regular calibration when home sphygmomanometers are used for therapeutic self-management of hypertension. |
| |
Keywords: | |
|