Replace内、外连接方式种植体-基台内部的应力分析
引用本文:Lan J,Xu X,Jiang G,Guan Y,Huang H,Liang J,Ma X. Replace内、外连接方式种植体-基台内部的应力分析[J]. 华西口腔医学杂志, 2011, 29(5): 464-468. DOI: 10.3969/j.issn.1000-1182.2011.05.004
作者姓名:Lan J  Xu X  Jiang G  Guan Y  Huang H  Liang J  Ma X
作者单位:1.山东大学口腔医院种植中心, 山东省口腔生物医学重点实验室;2.山东大学材料科学与工程学院, 济南250012
基金项目:国家“十五”科技攻关计划基金资助项目(2004BA720A-27)
摘    要:
目的 建立2种Replace实体种植体的下颌骨三维有限元模型,研究Replace内、外连接方式种植体-基台内部的应力状态.方法 测量2种种植体各部件的数据和利用螺旋CT扫描下颌骨截面形态,分别建立2种种植体的三维骨内模型,对模型采用轴向加载(200 N)、30°侧向加载(100N),分析2种种植体-基台内部的应力分布趋...

关 键 词:牙种植体  基台  有限元分析
收稿时间:2011-10-25
修稿时间:2011-10-25

Three-dimension finite element analyses of interior stress of two kinds of Replace implant
Lan Jing,Xu Xin,Jiang Guangshui,Guan Yanjin,Huang Haiyun,Liang Jin,Ma Xiaoni. Three-dimension finite element analyses of interior stress of two kinds of Replace implant[J]. West China journal of stomatology, 2011, 29(5): 464-468. DOI: 10.3969/j.issn.1000-1182.2011.05.004
Authors:Lan Jing  Xu Xin  Jiang Guangshui  Guan Yanjin  Huang Haiyun  Liang Jin  Ma Xiaoni
Affiliation:1. Implantation Center, School of Stomatology, Shandong University, Shandong Provincial Key Laboratory of Oral Biomedicine, Jinan 250012, China; 2. School of Materials Science and Engineering, Shandong University, Jinan 250012, China
Abstract:
Objective To establish three-dimension finite element model of mandible with two kinds of Replace implant, and to study stress of implant and abutment. Methods The data of components of the dental implant was measured, cross section of the mandible was scanned by spiral CT and image reconstruction technique was conducted. Threedimension finite element analysis software UG and MSC. Marc/Mentat were used to built the conjunction model and bone model of two implant systems. Axial loading(200 N) and 30° oblique loading(100 N)were applied on the models respectively, the stress distribution patterns of the implant and abutment of two implant systems were analyzed. Results The peak stress district was concentrated on the structure of the implant cervix, which was more obviously displayed on the Replace Select implant. The peak stress of oblique loading was higher than that of axial loading. The peak stress on the implant cervix of Replace Select implant was higher than that of Replace External Hex implant in all loadings. Conclusion To Replace Select especially, oblique force should be avoided on clinical practice in case of the implant fracture.
Keywords:dental implant  abutment  finite element analysis  